CM-36d
The Spectrophotometer CM-36d is an entry-level benchtop color measuring instrument with advanced functions that provide cost-effective solutions to today’s measurement needs.
With an inter-instrument agreement (IIA) of ΔE*ab< 0.15 and repeatability of σ∆E*ab < 0.03, the CM-36d can assure the measurement and its data are communicated within the supply chain consistently.
The CM-36d is design with enhanced useability. The instrument comes with a flat-back design that allows its orientation to be repositioned to take measurements vertically (top-port concept). The front plate of the CM-36d comes with four screw holes that enable easy installation of custom adapter/sample holders. It also has three different aperture sizes (Ø 4.0 mm, 8.0 mm, and 25.4 mm) for different measurement needs.
With an easy-to-read status panel with the measuring button, the CM-36d can displays the measurement status and condition settings at a glance, reducing operator error and improve work efficiency. Accurate positioning of the measurement sample is made easy with its sample viewing function that utilizes an integrated camera to provides a clear preview of the sample.
The CM-36d offers worry-free and high-reliability measurement through its optional WAA (Wavelength Analysis & Adjustment) function that analyzes and adjusts for any wavelength shift during each calibration. This innovative function ensures the instrument and measurement are stable and reliable over time.
Color | Illumination/ Viewing System | Reflectance:
di:8°, de:8° (diffused illumination, 8° viewing), SCI (Specular Component Included) / SCE (Specular Component Excluded) switchable. Conforms to CIE No.15, lSO7724/1, ASTM E1164, DIN 5033 Teil7, and JIS Z 8722 condition c standards |
Integrating Sphere Size | Ø152 mm (6 inches) | |
Detector | Dual 40-element silicon photodiode arrays | |
Spectral Separation Device | Diffraction grating | |
Wavelength Range | 360 to 740 nm | |
Wavelength Pitch | 10 nm | |
Half Bandwidth | approx. 10 nm | |
Reflectance
Range |
0 to 200% | |
Resolution: 0.01% | ||
Light Source | 1 Pulsed xenon lamps | |
Illumination / Measurement Area | LAV: Ø30.0 mm / Ø25.4 mm
MAV: Ø11.0 mm / Ø8.0 mm SAV: Ø7.0 mm / Ø4.0 mm |
|
Repeatability | Colorimetric values: Standard deviation within ΔE*ab 0.03
Spectral reflectance: Standard deviation within 0.1% (When a white calibration plate is measured 30 times at 10-second intervals after white calibration) |
|
Inter-Instrument Agreement | Within ΔE*ab 0.15
(Based on average for 12 BCRA Series II color tiles; LAV/SCI. Compared to values measured with master body under Konica Minolta standard measurement conditions) |
|
UV Setting | No adjustment function (UV 100%) | |
Measurement Time | Approx. 3.5 second (SCI+SCE measurement) | |
Minimum Interval Between Measurements | Approx. 4 second (SCI+SCE measurement) | |
Sample Viewer Function | Using internal camera
(Image viewable/copiable using optional software such as SpectraMagic NX software Ver. 3.2 or later) |
|
Internal Performance Check*2 | WAA (Wavelength Analysis & Adjustment) Technology | |
Interface | USB 2.0 | |
Target Mask Auto Detection | Yes | |
Power | Dedicated AC adapter | |
Operating temperature / humidity range | 13 to 33°C, Relative humidity: 80% or less (at 35°C) with no condensation | |
Storage temperature / humidity range | 0 to 40°C, Relative humidity: 80% or less (at 35°C) with no condensation | |
Size (W x H xD) | Approx. 248 x 250 x 498 mm | |
Weight | Approx. 8.3 kgKONICA MINOLTA |